O|1|000663|36^0044^2^^SAMPLE^NORMAL|ALL|R|20050705093416|||||X||||||||||||||O [CR]
(1)(2) (3) (4) (5) (6) (7) (8) (11)
* 4,*6 *5 *3
O|1|000663|36^0044^2^^SAMPLE^NORMAL|^^^10^2|R|20050705093416|||||N||||||||||||||Q [CR]
(1)(2) (3) (4) (5) (6) (7) (8) (11)
* 4 *5 *2
O|1|000663|6^44^2^^S1^SC|^^^10^2|R||20050705093416||||N||||1|||||||20050705095504|||F[CR]
(1)(2) (3) (4) (5) (6) (7) (8) (9) (10) (11)
O|1|000663|6^44^2^^S1^SC|^^^10^2|R||20050705093416||||A||||1||||||||||O[CR]
(1)(2) (3) (4) (5) (6) (7) (8) (9) (11)
Pos No Field Type Max EV CV
Comments
Elecsys Type Format
cobas Type Format
1 (1) Record Type ID ST 1 X R O is fixed.
2 (2) Sequence Number NM 6 X R Indicates sequence No. Normally it is 1.
3 Specimen ID ST 22 X R Indicates sample No. (Sample
ID.)
For control sample, set name
of control.
Indicate sample No. (Sample
ID.)
When barcode read error
occurs, the sample id will be
generated as follows:
@<Sequence no.>
For control sample, the control
name or control name^lot
number is sent to the host,
according to the settings at
the host communication
setting screen.