cobas e 411 analyzer
Roche Diagnostics
Host Interface Manual for use in the US
Document Version 2.3 55 of
61
1907-05-0816
Batch test selection information
Download test selection information of sample ID=000051 from HOST.
H|\^&|||host^1|||||cobas-e411|TSDWN^BATCH|P|1[CR]
P|1 [CR]
O|1|000051|^^^^S1^SC|^^^10^/^^^30^2/^^^40^|R||||||A||||1
||||||||||O[CR]
L|1|N[CR]
Download of test selection information is new order only. Sequence No. is automatically
numbered by the cobas e 411 analyzer.
The cobas e 411 analyzer can store 2000 tests, e.g., for 1000 samples with two tests. The older
tests are overwritten with a new order if 2000 tests are already stored.