Data alarm list
Flag Data Alarm Screen/Printer Elecsys Type Host No. cobas Type Host No.
1 Power Fail/Power Off Cancel Cancel 1 -
2 E.STOP Cancel Cancel 2 -
3 STOP Cancel Cancel 3 -
4 P.STOP/A.STOP Cancel Cancel 4 -
5 S.STOP Cancel Cancel 5 -
6 Recovery Cancel Cancel 6 -
7 Sample Short Samp.S 7 3
8 Assay Reagent Short Reag.S 8 4
9 Diluent Short Reag.S 9 4
10 Pretreatment Reagent Short Reag.S 10 4
11 <Not available> - - -
12 Abnormal Reagent Disk Temperature Reag.T 12 74
13 Abnormal Incubator Temperature Inc.T 13 75
14 Abnormal Measuring Cell Temperature Cell.T 14 77
15 Abnormal System Reagent Temperature SysR.T 15 76
16 System Reagent Short SysR.S 16 62
17 ADC abnormal ADC.E 17 1
18 <Not available> - - -
19 <Not available> - - -
20 <Not available> - - -
21 <Not available> - - -
22 <Not available> - - -
23 <Not available> - - -
24 Calculation Error Calc.? 24 39
25 No Calibration Data Cal.E 25 43
26 Previous Calibration Data Cal.E 26 43
27 <Not available> - - -
28 <Not available> - - -
29 <Not available> - - -
30 <Not available> - - -
31 Assay Reagent Hovering Reag.H 31 69
32 Diluent Hovering Reag.H 32 69
33 Pretreatment Hovering Reag.H 33 69
34 <Not available> - - -
35 Assay Reagent Film Detected Reag.F 35 70
36 Diluent Film Detected Reag.F 36 70
37 Pretreatment Film Detected Reag.F 37 70
38 System Reagent Film Detected Reag.F 38 70
39 <Not available> - - -
40 AB Level Range Over >AB 40 63
41 AB Level Check Error AB.E 41 64
42 Current Range Over >Curr 42 65
43 Current Range Check Error Curr.E 43 66
44 System Reagent Temperature Unstable SysR.U 44 120
45 Sample Clot Detected Samp.C 45 72