EasyManuals Logo

Roche cobas e 411 User Manual

Roche cobas e 411
61 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #51 background imageLoading...
Page #51 background image
cobas e 411 analyzer
Roche Diagnostics
Host Interface Manual for use in the US
Document Version 2.3 51 of
61
1907-05-0816
The cobas e 411 analyzer sends test selection information such as sample
ID=000004, sequence No.=40, Disk No=0, and Position=5.
H|\^&|||cobas-e411^1|||||host|TSREQ^REAL|P|1[CR]
Q|1|^^000004^40^0^5^^S1^SC||ALL||||||||O[CR]
L|1|N[CR]
Example 1: HOST replies test selection information of sample ID=000004.
H|\^&|||host^1|||||cobas-e411|TSDWN^REPLY|P|1[CR]
P|1 [CR]
O|1|000004|40^0^5^^S1^SC|^^^10^\^^^30^2\^^^40^|R||||||A
||||1||||||||||O[CR]
L|1|N[CR]
Example 2: HOST replies without order information of sample ID=000004 or
the generated sample id=@<sequence no.>.
H|\^&|||host^1|||||cobas-e411|TSDWN^REPLY|P|1[CR]
P|1 [CR]
O|1|000004|40^0^5^^S1^SC||R||||||A||||1||||||||||O[CR]
L|1|N[CR
The cobas e 411 analyzer sends a cancel to HOST when HOST does not reply
within 15 seconds.
H|\^&|||cobas-e411^1|||||host|TSREQ^REAL|P|1[CR]
Q|1|^^000004^40^0^5^^S1^SC||ALL||||||||A[CR]
L|1|N[CR]

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Roche cobas e 411 and is the answer not in the manual?

Roche cobas e 411 Specifications

General IconGeneral
BrandRoche
Modelcobas e 411
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals