cobas e 411 analyzer
Roche Diagnostics
40
of 61 Host Interface Manual for use in the US
Document Version 2.3
1907-05-0816
Pos No Field Type Max EV CV
Comments
Elecsys Type Format
cobas Type Format
12 (8) Action Code ST 2/1 X R Indicates current status of
sample (*3)
X: measured (upload)
N: new sample order (download)
X\Q: control sample (upload)
Q: control sample (download)
A: additional test order (upload)
Indicates type of information to
be reported (*3)
N: communication of patient
sample result from analyzer.
(upload)
Q: communication of control
sample result from analyzer
(upload)
A: test order from HOST.
(download)
13 Danger Code
14 Relevant Clinical
Information
15
Date/Time Specimen
Received
16 (9) Specimen Descriptor NM 1 R Indicates sample type (*2)
1: blood serum.
2: urine.
5: others.
17 Ordering Physician
18
Physician’s Telephone
Number
19 User Field No. 1
20 User Field No. 2
21 Laboratory Field No. 1
22 Laboratory Field No. 2
23
(10)
Data/Time Results
Reported or Last
Modified
TS
14
X
Indicates date when all test
results are collected.
Setting from HOST is not
applicable. Setting is as
follows:
YYYYMMDDHHMMSS
24
Instrument Charge to
Computer System
25 Instrument Section ID
26 (11) Report Types ST 1 X R Indicates report type.
Q: response to inquiry. (download)
Z: no response request to inquiry.
(download)
O: from e411 to HOST. (upload)
Indicates type of
communication.
O: test order. (download)
F: communication of result.
(upload)
27 Reserved Field
28
Location or Ward of
Specimen Collection
29
Nosocomial Infection
Flag
30 Specimen Service
31 Specimen Institution
(*1) MC indicates reduced volume not micro cup (like the cobas 6000 analyzer series).
(*2) Except for quality controls, the Specimen Descriptor of Instrument Specimen ID for cobas type is prior to Sample Type when they are mismatched.
(*3) Action Code of already ordered sample for Elecsys type is N: When new order of sample is received, delete existing test information to order the receiving tests.
Same for cobas type as its Action Code is "A" fixed (download).
(*4) Application Code in Universal Test ID.
(*5) See Table 17 for automatic dilution factor/ratio for <Dilution> of Universal Test ID.
(*6) Sending of multiple orders for the same test in one sample record is not allowed.
Table 16: Test order record