Setting Up the Instrument
R&S
®
RTO2000
101Getting Started 1332.9719.02 ─ 08
Two connector pins are located on the front panel. The right pin is on ground level. The
left pin supplies a square wave signal with 1 kHz for low frequency probe compensa-
tion.
1. Connect the BNC connector of the probe to input [CH1].
2. Connect the probe's ground connector to the right compensation pin, and the tip
with the left pin.
3. Press [Autoset].
A square wave appears on the display.
4. Adjust the compensation trimmer of the probe to optimum square wave response.
For details, refer to the documentation of your probe.