EasyManua.ls Logo

Rohde & Schwarz R&S RTO2000 Series - Performing a Mask Test

Rohde & Schwarz R&S RTO2000 Series
106 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Trying Out the Instrument
R&S
®
RTO2000
65Getting Started 1332.9719.02 ─ 08
4.11 Performing a Mask Test
In the following example, you perform a mask test to determine whether the signal
exceeds a rectangular area.
The usage of masks tests is also shown in a short video that is available on the instru-
ment: "File" menu > "Tutorials" > "Mask Test".
1. To restore the default signal channel settings, press [Preset] and [Autoset].
2. Tap the "Masks" icon on the toolbar.
3. Tap the corner points of the first mask segment on the touchscreen. Select the cor-
ner points of one complete pulse, with a minor offset to the inside.
Tip: To create an exact rectangle, draw the diagonal of the rectangle on the
screen.
4. Tap "Finish segment" in the sidebar.
5. Add another mask segment above the positive pulse:
a) Tap "Add segment" in the sidebar.
b) Tap two points slightly above the pulse.
Performing a Mask Test