Trying Out the Instrument
R&S
®
RTO2000
51Getting Started 1332.9719.02 ─ 08
6. To remove the result icon and make room on the display for other results, tap the
red cross on the icon label.
4.7.2 Performing an Amplitude Measurement
To measure the voltages of a pulse quickly, measure the amplitude in the time domain.
1. To restore the default signal channel settings, press the [Preset] and [Autoset]
keys.
2. Tap the "Measurement" icon on the toolbar.
3. In the sidebar, select the measurement types, for example, "High", "Low", and
"Amplitude".
4. Tap the diagram in which you want to measure.
The selected measurement types are enabled, using the selected (focused) wave-
form as the source. The "Measurements" result box shows the measured values.