Trying Out the Instrument
R&S
®
RTO2000
37Getting Started 1332.9719.02 ─ 08
8. Tap the "Undo" icon on the toolbar repeatedly until the trigger mode is "Auto" and
the trigger source is "CH1".
9. Press the [Slope] key to toggle the trigger slope.
Watch the waveform and the "Trigger" settings.
4.2 Acquiring Data
Although you are using only one input channel, you can acquire the calibration data
using three different arithmetic methods and display the results in separate waveforms.
1. In the [Horizontal] area, press the [Acquisition] key.
2. In the "Acquisition" tab of the "Horizontal" dialog box, enable all three waveforms
("Wfm1", "Wfm2", "Wfm3").
If you cannot see these settings, enable "Multi waveform" in the lower right corner
of the dialog box.
3. For Wfm1, select the "Mode" = Sample and "Wfm Arithmetic" = "Off".
4. For Wfm2, select the "Wfm Arithmetic" Envelope. The "Decimation" type is auto-
matically set to Peak detect to display the correct envelope waveform.
5. For Wfm3, select the "mode" = High Res and the "Wfm Arithmetic" to Average.
Acquiring Data