Trying Out the Instrument
R&S
®
RTO2000
36Getting Started 1332.9719.02 ─ 08
4. If necessary, compensate the passive probe as described in Chapter 6.4, "Adjust-
ing Passive Probes", on page 100.
5. In the [Trigger] area of the front panel, press the [Source] key. Press the key again
to switch the trigger source to "C2".
An unstable waveform is displayed. In auto mode, the instrument triggers repeat-
edly after a time interval if no real trigger occurs.
6. In the [Trigger] area, press the [Mode] key.
7. Check the "Trigger" settings in the upper right corner of the screen.
The trigger mode has changed to "Normal". The waveform is no longer refreshed,
and "Wait" is displayed in the trigger settings. The instrument cannot find a real
trigger event because there is no signal on channel 2.
Displaying a Basic Signal