Trying Out the Instrument
R&S
®
RTO2000
66Getting Started 1332.9719.02 ─ 08
c) Tap "Finish segment" in the sidebar.
d)
Tap   in the result box.
e) Select the "Mask Definition" tab.
f) In the "Region" column of the second mask segment, select "Upper".
6. Select the "Test Definition" tab.
7. Select channel 1 as the "Source".
8. Define the number of tolerable sample hits in the "Violation tolerance" field.
A test has failed if the number of sample hits exceeds the limit of "Violation toler-
ance".
9. Select the "Event Action / Reset" tab.
10. For the "Stop acquisition" action, select "On violation".
When the violation tolerance is exceeded, the acquisition stops. The results of the
mask test are shown in the "MaskTest" results box. Mask hits are also indicated as
red points in the mask segment in the diagram.
Performing a Mask Test