Preparing for use
R&S
®
SMB100B
25Getting Started 1422.1045.02 ─ 06
2
1
1 = Hostname
2 = IP address
If disconnected from the LAN, the R&S SMB100B displays the IP address
0.0.0.0.
See also chapter "To configure the instrument for remote access" in the user
manual.
4.8 Connecting USB devices
You can connect or disconnect all USB devices from the R&S SMB100B during
operation.
To connect USB storage devices
USB storage devices, such as memory sticks, allow data transfer from or to the
R&S SMB100B. You can also use them for firmware updates.
► Connect the USB storage device to any of the USB connectors.
To connect USB devices with external power supply
1. NOTICE! Connected devices with external power supply can feed back cur-
rent into the 5 V power supply of the USB interface and thus damage the
R&S SMB100B.
Ensure that there is no connection between the positive pole of the power
supply and the +5 V power pin of the USB interface (VBUS).
2. Connect the USB storage device to any of the USB connectors on the front
panel or rear panel.
To connect a keyboard
► Connect the keyboard to any of the USB connectors on the front panel or rear
panel.
Connecting USB devices