4.12 Switching on or off
The following table provides an overview of power states, LEDs and positions of
the power switch.
Table 4-2: Overview of power states
State LED Position of power switch
Off
gray
[0]
Standby
orange
orange
[I]
Ready
yellow
[I]
To switch on the R&S SMB100B
The R&S SMB100B is off but connected to power. See Chapter 4.6, "Connecting
to power", on page 24.
1. Set the switch on the power supply to position [I] on the rear panel.
The LED of the [On/Standby] key on the front panel is orange.
2. Wait until the oven-controlled oscillator (OCXO) warms up. For the warm-up
time, see the data sheet.
3. Press the [On/Standby] key on the front panel.
The LED changes to green. The R&S SMB100B boots.
After booting, the R&S SMB100B starts up displaying the home screen on the
screen on the front panel.
To check start-up functions
When starting for the first time, the R&S SMB100B starts with the default settings.
When restarting, the R&S SMB100B recalls the instrument configuration before
shut-down.
See chapter "Saving and recalling instrument settings" in the user manual.
When switched on, the R&S SMW automatically monitors main functions. You
can query erroneous functions. In addition to automatic monitoring, you can per-
form maintenance tasks. See the following:
●
Chapter "Querying notifications" in the user manual.
●
Chapter "Performing maintenance tasks" in the user manual.