Never use meter if the meter or test leads are damaged or broken.
Never use uncased meter.
Be sure to use a fuse of the specified rating or type. Never use
a substitute of the fuse or never make a short circuit of the fuse.
Always keep your fingers behind the finger guards on the
probe when making measurements.
Be sure to disconnect the test pins from the circuit when
changing the function or range.
Never use meter with wet hands or in a damp environment.
Never open tester case except when replacing batteries or
fuse. Do not attempt any alteration of original specifications.
To ensure safety and maintain accuracy, calibrate and check
the tester at least once a year.
Indoor use.
Correct measurement may not be performed when using the
meter in the ferromagnetic / intense electric field such as
places near a transformer, a high-current circuit, and a radio.
The meter may malfunction or correct measurement may not
be performed when measuring special waveform such as that
of the inverter circuit.
CAUTION
5.
6.
7.
8.
9.
10.
11.
12.
13.
1.
2.
Maximum rating
input value
Function Input
Maximum overload
protection value
1-3 Maximum Overload Protection Input
DCV
ACVHzDuty
Ω
/ /
V,Hz,
Duty,Ω,
μA, mA
(+)
and
COM
(−)
,
,
615 A(+)
and
COM(−)
DC 1000 V
AC 750 V
Voltage and current
input prohibited
DC/AC 600 mA,
Voltage input
prohibited.
DC/AC 15 A,
Voltage input
prohibited.
DC
AC 6
15 A
DCAC 6006000 μA
DC
AC 60600 mA
DC 1000 V,AC 750 V
or peak max. 1000 V
0.4 A/1000 V fuse*
Breaking capacity 30 kA
16 A/1000 V fuse
Breaking capacity 30 kA
* According to the time-current characteristic curve of the 0.4 A fuse,
it will never blow at 0.6 A but blows in less than 0.1 sec. when the
current exceeds 1.5 A. This characteristic is completely suitable for
use with this instrument.