42
– Enter a value for the local elevation at the place of installation in meters above
sea level (reference value: 151) (key sequence: ) k k).
Note: A negative value may be entered for local elevation; for example,
if the scale is installed in a mine.
– Press ) to confirm your input.
– Activate menu item 1-20-4 to store new data
(key sequence: ) k k).
Rather than the geographical latitude and local elevation of the place of
installation, you can enter a value for the acceleration of gravity at the place of
installation (menu item 1-20-3).
Example: Entering the acceleration of gravity place of installation.
– Select menu item 1-20-3 (acceleration of gravity) (key sequence: k k )).
Example: In the example shown here, the geographical latitude and local
altitude (menu items 1-20-1 and 1-20-2) were the last parameters
entered in menu line 1-20. Thus the input field for acceleration of
gravity is filled with 000…
In the example shown here, the most recently entered parameter
was the value for the acceleration of gravity at Goettingen, Germany
(menu item 1-20-3). This value is the reference value, which means
no correction of the span adjustment for the place of installation
is required.
display_000.eps
display_981.eps
display_151.eps