43
– Enter the acceleration of gravity at the place of installation (in ms
-2
/ or the
reference value: 9.811590). Press ) to confirm your input.
– Activate menu item 1-20-4 to store new data (key sequence: ) k).
The display shows “busy”, indicating that the parameters entered are being stored.
– Restart the scale: Turn the indicator off and then on again (key sequence: e
e, or press and hold the ) key (> 2 sec).
After the self-test of the display, the scale is ready to operate.
Note: Once the precise geographical data or acceleration of gravity has been entered, the
tolerance range is restricted (see page 23).
– After data has been entered as described above, the span must
be adjusted.
Example
Settings (changes in the factory settings required for this example):
Menu item 1-9-1: External calibration/adjustment with default weight;
Menu item 1-10-1: Calibration/adjustment in one operation;
Menu item 1-16-1: External calibration/adjustment/linearization
accessible.