39
Example
Fill containers to approximately identical weights (± 30 %)
Configuration (changes in the factory settings required for this example):
Menu path: Application > Second tare memory with auto-taring (2 1 10)
Menu path: Weight unit 2 > Grams (3 1 2)
Step Key (or instruction) Display/Output
1. Turn on the scale e
2. Zero the scale, if necessary ( 0.000 kg
3. Place empty container on + 0.065 kg
the scale (in this case, 65 g)
4. Save weight value in second v 0.000 kg
NET
tare memory
If the printout is configured to
to include data ID codes, the
data is printed as shown here N1    +    0.065 kg
5. Fill container (in this case, to + 0.500 kg
NET
500 g)
6. Display weight unit 2 w + 500 g
NET
7. Display weight unit 1 w + 0.500 kg
NET
8. Remove filled container – 0.065 kg
NET
9. Place next container on 0.000 kg
NET
 (the scale
scale (here: 75 g) is tared automatically)
10. Repeat procedure, starting
 from Step 5.