10
1.  Check that the battery protector is removed and 
the battery cover is properly closed.
2.  Place the pipette into the charging unit. Make 
sure that the metal contacts at the rear of the 
pipette are properly fitted to the tracks of the 
metal contacts in the top of the charging unit.
3.  When charging, the battery symbol will be 
displayed with the + and – symbols flashing. The 
charging time for completely discharged  batteries 
is 4 hours.
4.  When the battery is full the battery symbol 
disappears from the display. If display shows 
RESET and PRESS TIP, press the tip ejection button 
to reset the pipette, the pipette will then be ready 
for use.
NOTE: Never reset the pipette with the tip attached 
as the attached tip can induce an error in the home 
position determination of the piston and tip ejector 
collar!
NOTE: When the pipette is not in use it is recommended 
to keep the pipette in the charging unit. This ensures 
that the battery is kept charged and the pipette is 
ready for use.
5.2.  Charging Directly
The AC-adaptor supplied with the eLINE® pipette or 
single charging stand can be used for the charging of 
the eLINE® electronic pipette. The socket for the AC-
adaptor is situated on the back of the pipette.
1.  Before connecting the AC-adaptor to the electrical 
outlet, check that its voltage settings corresponds 
to the voltage settings of your mains electrical 
outlet.
2.  Check that the battery protector is removed and 
the battery cover is properly closed.
3.  Connect the AC-adaptor to the mains electrical 
outlet as well as to the socket at the back of the 
pipette.
4.  The charging time is 12 hours for a completely 
discharged battery.