27
NOTE:  This procedure overwrites the possible 
previously stored mode from the selected storage 
location.
9.2.  Recall stored mode from the 
storage location
To recall a stored mode from Prog 1–6:
1. Press 
 to display PROG
2.  Scroll with 
 until the desired storage location 
is displayed (1–6)
NOTE: The display shows simultaneously the mode 
symbol of the operating mode stored in to the storage 
location in question.
3.  Confirm by pressing 
4.  The pipette is ready to perform according to the 
stored mode. Position the tip to aspirate and press 
START button 
.
10. Pipetting  
Recommendations
The wide selection of different operating modes 
makes it possible to perform several different liquid 
handling tasks with eLINE® pipettes. The operating 
modes P, P+MIXING, P COUNT, dd, dd+MIXING and 
SA feature an automatic blow-out and others leave 
an excess liquid in the tip. The user should follow 
the recommendations below to ensure optimal 
performance.
10.1.  Dispensing With Blow-out
The automatic blow-out function, in different 
pipetting and diluting modes, is followed by an 
immediate return of the piston to the home position. 
To avoid accidental aspiration of the liquid back into 
the tip, it is recommended that the dispensing is 
always performed above the liquid surface.