8.4. Multi-aspirating (SA)
The pipette performs consecutive aspiration
operations of the programmed volume. This mode
allows e.g. emptying a microwell plate, sample pooling
and other special applications.
1. Press
to display MULTI
2. Scroll with
until the symbol SA is displayed.
3. Confirm by pressing
4. Select the desired aspirating volume by using
NOTE: When holding down the volume display
starts to scroll faster.
5. Press
or to confirm the selection
6. Select the desired number of aspirations with
7. Press to confirm selection and to display the
speed in or press
and continue from Step 12.
8. Select the speed in with
(1= Min and 9 = Max)
9. Press
or to confirm selection and to
display the speed out
10. Select the speed out with
(1= Min and 9 =
Max)
11. Press
or to confirm selections
12. Position the tips to aspirate and press the START
button
repeating the action until the cycle is
complete. The total volume is displayed.
13. Position the tips to dispense and press START
button twice
.
8.5. Automatic dispensing (Ad)
The pipette performs repetitive dispensing of a selected
volume automatically at the specified interval (0.1–9.9
seconds). The sum of the dispensing aliquots and an
automatically selected excess volume is aspirated into
the tip. The excess volume is needed to ensure equal
operating conditions for each dispensing step.