16
To mix:
1.
Position the tip in to the solution, then press and
hold the START button
.
Mixing is performed automatically as long as the
START button
is held down.
2. Position the tip to dispense. Press START button
twice
NOTE: The mixing is performed with about 70% of the
total volume.
7.2. Reverse Pipetting (rP)
A selected volume plus and excess is aspirated into the
tip. After dispensing the selected volume the excess
volume remains in the tip and is discarded.
1. Press
to display PIPET
2. Scroll with
until the symbol rP is displayed.
3. Confirm by pressing
4. Select the desired pipetting volume by using
NOTE: When holding down the volume display
starts to scroll faster.
5. Press
to confirm selection and to display the
speed in
or press
and continue from Step 10.
6. Select the speed in with
(1= Min and 9 = Max)
7. Press
or to confirm selection and to
display the speed out
8. Select the speed out with
(1= Min and 9 =
Max)
9. Press
or to confirm selections
10. Position the tip to aspirate and press the START
button
11. Position the tip to dispense. Press the START
button