NOTE: After recalibration the display will show the 
operating mode symbol followed with letters CAL to 
indicate that the mode and the volume in question is 
recalibrated to give out the new volume.
NOTE: Changing the mode will reset the recalibration. 
However, it is possible to store the recalibrated mode 
to the storage location in PROG mode as described in 
paragraph 8.
13. Troubleshooting
The  eLINE® pipettes have a onboard monitoring 
program to control the performance of each pipetting 
action. If any error message appears to the display it 
means that the pipette has been unable to perform 
the attempted action properly. In case of an error the 
display will show Error and PRESS  Ent. message. To 
clear the message please follow these steps:
1.  As this procedure will reset the pipette and the tip 
ejector please first remove the tip by hand.
2.  Recharge the pipette for 15 minutes.
3.  Clear the error message by pressing 
. The 
display will show RESET and PRESS TIP. 
3.  Reset the pipette by pressing either of the tip 
ejector buttons 
.
NOTE: Repeated occurrence of error messages 
indicates an internal error causing failure of the 
eLINE® to complete the execution of the pipetting. The 
eLINE® therefore needs to be returned to the nearest 
Sartorius Service Centre or your local Sartorius service 
representative for repair.