17
12. Position the tip to discard excess liquid and press
the START button twice
NOTE: It is also possible to continue to aspirate and
dispense the same volume without the empty function.
To continue, keep the START button depressed during
dispensing and within one second the direction arrow
will change. Continuing to hold the START button
down, place the tip into the liquid again and then
release the START button to aspirate the next sample.
7.3. Pipetting With Cycle Counter
(P COUNT)
The eLINE® performs pipetting with blow-out and
the cycle counter function. The cycle counter counts
up to 99 pipettings and then returns back to 0. It is
possible to choose the cycle number to begin with. It is
also possible to change the cycle number or reset the
counter during the pipetting by using the arrow keys.
1. Press
to display PIPET
2. Scroll with
until the symbol P and COUNT is
displayed.
3. Confirm by pressing
4. Select the desired pipetting volume by using the
NOTE: When holding down the volume display
starts to scroll faster.
5. Press
or to confirm selection.
6. Cycle counter will show 1. Use
to select the
sample number you want to begin with (0-99).
7. Press
to confirm the selection and to display
the speed in
or press
and continue from Step 12.
8. Select the speed in with
(1= Min and 9 = Max)
9. Press
or to confirm selection and to
display the speed out