If the sample is liquid containing solids, use one of the following tools:
 – Glass stirrer
 – Spoon
 – Magnetic stirrer
 t Use an appropriate tool for grinding the sample.
Use a disposable sample pan:
 t Use only Sartorius disposable sample pans (inner diameter of 90 mm). 
Reusing sample pans leads to poor repeatability of results:
 – After cleaning, sample residues can still remain on the pan.
 – Residues from cleaning agents can evaporate during the next moisture 
analysis. 
 – Scratches and grooves made during cleaning result in damage that forms 
points of attack for hot, rising air produced during the drying process 
(more pronounced buoyancy).
16
Operating Instructions Moisture Analyzer MA160
  Basic Principles