59
Calibration and Adjustment
Operating Instructions Moisture Analyzer MA160
Viewing the Calibration Report
Calibration/Adjustment
MA performance test
Weighing system adjustment
Calibration report
Cal./Adj. the Heating Module
t To view the most recent calibration reports: Go to the Calibration/Adjustment
menu and select Calibration report.
You can select recent reports (see following figure) and then sort the calibration
reports by date. Up to 99 entries per day are saved. Calibration reports older
than 30 days are deleted. Only the current calibration report is saved for each
calibration or adjustment of the heater.
y ”Calibration report” menu.
Example of a Calibration Report:
--------------------
2016-03-29 12:02
Sartorius
Mod. MA160-1
SerNo. 0031710609
BAC: 00-53-01
APC: 01-80-01
MCU: 07-40-05
--------------------
2016-01-30 12:02
External calibration
Start: manually
Set + 100.000 g
Dev - 0.002 g
External adjustment
Dev 0.000 g
Temp + 23.7 ’C
--------------------
Name:
--------------------
Calibration Report
MA Performance Test
Weighing system adjustm.
Heater calibration
Heater adjustment