MA Performance Test
Operating Instructions Moisture Analyzer MA160
MA Performance Test
The performance test can be used to check whether the analyzer has a fault or is
ready for operation. During a performance test, both the heating unit and the
weighing system are tested at the same time. A ReproEasy pad (accessory) is
required to conduct a performance test.
Starting the Performance Test
Calibration/Adjustment
MA performance test
Weighing system adjustment
Heater module cal./adj.
Calibration report
t Select “MA performance test.”
y The display switches to a view with three heating temperatures.
Ensure that the analyzer has not been heated for at least one hour before
conducting a performance test.
MA performance test
The device is too warm for a performance
test.
The device is still cooling down.
m
t Select < to confirm the message and start the performance test again later.
MA Performance Te st
Temperature 85 °C
Temperature 125 °C
Temperature 165 °C
t Select one of the suggested temperatures given for the method you are using.
Open hood
0.000 g
D
t Open the hood.
y The next step is displayed when the hood is opened.