58
Calibration and Adjustment
Operating Instructions Moisture Analyzer MA160
Settings
Language
Date and time
Device information
Calibration/Adjustment
y The Settings (Setup) menu appears.
t Select Calibration/Adjustment.
Calibration/Adjustment
MA performance test
Weighing system adjustment
Calibration report
Cal./Adj. the Heating Module
t Select Weighing system adjustment.
t Select CAL-Extern to calibrate the analyzer using an external calibration
weight.
y The Calibration/Adjustment window appears.
t Close the hood and select l to confirm.
y The CAL-Extern window appears.
t Zero the integrated weighing system if necessary.
y A message appears on the display, prompting you to place the calibration
weight on the analyzer (100 g).
t Place the calibration weight on the pan support.
y Calibration starts automatically.
When calibration is complete, a report appears on the display.
The report gives the deviation detected during calibration. You are asked
whether the weighing system should be adjusted. After adjustment the report
shows the result of the adjustment.
t Close the report: Select <.
y The moisture analyzer is now calibrated. If you have also carried out an
adjustment after calibration, the moisture analyzer is also now adjusted.
Calibrating/Adjusting the Heating Module
t To calibrate and adjust the heating module, please read the operating
instructions for the “YTM15MA heating adjustment set.”
CAL-Extern
START
Please set balance to zero
G - 0.401 g
CAL-Extern
Please place on predefined weight.
Close hood
Set - 100,000 g