77
Specifications
Operating Instructions Moisture Analyzer MA160
Specifications
Dryer Functions MA160-1
Temperature range and settings 40–200°C, in increments of 1°C; stand-by temperature selectable from 50–120°C
Sample heating
Infrared heating using AURI heater, 600 W
Heating programs Standard drying, gentle drying, MA35 mode
Switch-off criterion Options:
– Fully automatic
– Semi-automatic mg (1–50 mg/5–300 s)
– Semi-automatic % (0.1–5.0%/5–300 s)
– Time (2:00–99.59 min.)
– Manual
Result display Moisture in % M and g / dry weight in %S and g / ATRO in %M/S
Memory capacity for methods 100
Net weight, approx. 6.3 kg
Weighing Functions
Max. weighing capacity 200 g
Repeatability, typical Starting at an initial weight from approx. 1 g: ± 0.2%; for initial weight from approx. 5 g: ± 0.05%
Readability 1 mg; 0.01%
Typical sample quantity 5–15 g
External adjustment weight
(min. accuracy class)
100 g (E2)
Disposable sample pan dimensions d 90 mm
Interface Mini USB: Automatic detection of Sartorius printers YDP30 and YDP40,
direct data transfer to Microsoft
®
Windows programs without additional software
SD card Type: SD and SDHC, capacity: up to 32 GB, file system: FAT32
Power Supply
Input voltage
100–120 V AC and 210–240 V AC (automatic voltage detection), ± 10%, 50/60 Hz
Protection class and standards
Protection class I as per EN 61010-1/IEC 61010-1; up to 3000 m above sea level; IP20 as per EN 60529/
IEC 60529
Power supply Using country-specific power cable from Sartorius only
Power consumption Max. 640 W (typically)
Ambient Conditions
The specifications apply under the following ambient conditions:
Environment For indoor use only
Ambient temperature +10°C to +30°C
Operational capability Operation guaranteed between +5°C and +40°C
Storage and shipping -10°C to +60°C
Altitude Up to 3000 m above sea level
Relative humidity 15% to 80% for temperatures up to 31°C non-condensing,
decreasing linearly up to 50% relative humidity at 40°C
Safety of Electrical Equipment
In accordance with EN 61010-1/IEC 61010-1
Safety requirements for electrical equipment for measurement, control and laboratory use –
Part 1: General requirements
Electromagnetic Compatibility
In accordance with EN 61326-1/IEC61326-1
Electrical equipment for measurement, control and laboratory use –
EMC requirements – Part 1: General requirements
Defined immunity to interference: Suitable for use in industrial areas
Limitation of emissions: Class B [suitable for use in residential areas and areas that are connected to a low voltage network
that (also) supplies residential buildings]. Therefore, the analyzer can be used in both areas.