55
MA Performance Test
Operating Instructions Moisture Analyzer MA160
MA performance test
Device is tested with reference sample.
Temperature: 85°C
Duration: 05:00 min
00:13 min
y The display shows the remaining time for the test.
The curve for temperature-related weight loss is analyzed during the test. to
determine whether the drying loss per time unit falls within the tolerance range.
If it does fall within the tolerance range, the analyzer’s function is assured. The
measure for this is the calculated MApi value (Moisture Analyzer performance
index).
Canceling the Performance Test
t To cancel the performance test, select <.
MA performance test
Measurement canceled
y The message “Measurement canceled” appears.
t Select l to confirm.
y The display switches back to the menu.
Performance Test Measurement Not Successfully
Completed
If a fault has been detected:
MA performance test
Device is not OK.
5 MApi << 10 MApi
Temperature 85°C
Please contact Sartorius service
y The message “Device is not OK.” appears.
t Select a to print the performance test result if necessary.
t Select l to confirm.
Contact the Sartorius Service Center if the message “Device is not OK.” appears.