EasyManua.ls Logo

Sartorius MA160 - Performance Test Successfully Completed

Sartorius MA160
84 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
56
MA Performance Test
Operating Instructions Moisture Analyzer MA160
Performance Test Successfully Completed
The analyzer is in working order and displays the message “Device is OK.” if the
MApi value falls within the following ranges:
Temperature setting Tolerance range
85 between 9 and 11
125 between 16 and 18
165 between 21 and 25
al
MA performance test
Device is OK.
Result 10 MApi
Temperature 85°C
3
y If the test is successful, the message “Device is OK.” first appears. Then the
analyzer indicates that the sample can be removed.
t Select a to print the performance test result if necessary.
Remove old sample
t Select l to confirm.
y You can remove the sample and return to the menu.