MA Performance Test
Operating Instructions Moisture Analyzer MA160
Performance Test Successfully Completed
The analyzer is in working order and displays the message “Device is OK.” if the 
MApi value falls within the following ranges:
Temperature setting  Tolerance range
85 between 9 and 11
125 between 16 and 18
165 between 21 and 25
al
MA performance test
Device is OK.
Result 10 MApi
Temperature 85°C
3
  y  If the test is successful, the message “Device is OK.” first appears. Then the 
analyzer indicates that the sample can be removed.
t Select a to print the performance test result if necessary. 
 
 
Remove old sample
  t Select l to confirm.
  y  You can remove the sample and return to the menu.