MA Performance Test
Operating Instructions Moisture Analyzer MA160
+ 2.500 g
Place pan, close hood for taring
D
t Place a sample pan on the analyzer.
t Close the hood.
y The moisture analyzer tares automatically.
Open hood
0.000 g
D
t Open the hood and remove the sample pan.
t Open the packaging of the ReproEasy pad (accessory) and remove the pad.
t Remove the protective wrapping.
t Affix the pad to the center of the sample pan, sticky side down.
Fill in reference sample and
close hood to start
;
- 0.000 g
t Place the sample pan with the “ReproEasy pad” (accessory) reference sample on
the pan support.
Close hood to start
;
+ 0.466 g
t Close the hood.
y The performance test starts automatically.
During the performance test, the status light on the analyzer flashes, if the
light has not been switched off in the menu.