EasyManua.ls Logo

Sartorius MA160 - Starting the Analysis; End of Analysis

Sartorius MA160
84 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Starting the Analysis
After closing the hood:
If the initial weight condition is met, the analyzer prompts you to close the
hood. The start weight is applied as soon as the sample chamber is closed and
the measured value is stable.
End of Analysis
Fully automatic
The analysis ends as soon as the weight loss per 24 s is below an automatically
detected threshold.
Semi-automatic, absolute
The analysis ends as soon as the weight loss per the set interval is lower than
a user-defined threshold (set in milligrams).
The weight loss is entered by the user.
Semi-automatic, percentage
The analysis ends as soon as the weight loss in percent is lower than a user-
defined threshold (set in percent).
The percentage is entered by the user.
Time
The analysis ends as soon as the specified time has elapsed.
Manual
Switch off using the END key. The maximum analysis duration is 24 hours.
Weight
Moisture (steam)
Autom. stopping
Time
19
Operating Instructions Moisture Analyzer MA160
Basic Principles

Related product manuals