33
System Settings (Setup Menu)
Operating Instructions Moisture Analyzer MA160
Under Decimal separator (also available when “PC spreadsheet format” has been
selected):
Decimal point (factory setting): The number value is transmitted with a decimal
point to the PC program (example: 99.963 g).
Decimal comma: The value is transmitted with a decimal comma to the PC
program (example: 99,963 g).
Under Keyboard emulation (also available when “PC spreadsheet format” or
“PC text format” has been selected):
Universal (Num Lock On) (factory setting): Data is sent as special keyboard
characters in ASCII format (ALT + number pad).
Requirement: Numbers lock must be turned on on the PC keyboard.
English (USA): Data is sent according to a keyboard set to English (USA).
Requirement: For PC applications, such as MS Excel, the keyboard must be
set to English (USA).
Further information can be found in chapter “USB Port.”
2) Connection to Sartorius Lab Printer via the Supplied Printer Cable
Under Device/Protocol:
t Select the desired setting for the connection to the printer.
YDP30/YDP40: The connected printer is detected automatically and the
connection is established.
Off: The connection to the printer is deactivated.
3) Connection to a Serial Printer or Another External Serial Device via a 9-pin
Serial Interface (USB Mini-A to RS-232)
Under Device/Protocol:
t Select the desired setting for the connection to the device.
Printer YDP20-0CE
USB-RS-232 SBI
Off: The connection is deactivated.
Check the settings on the Printer YDP20-0CE:
t Set the printer to factory settings:
1200 baud, 7 bits, ODD parity, 1 stopbit, CTS/RTS handshake
Under RS-232 Configuration:
This option can be selected if under Device/Protocol the setting USB-RS-232 SBI
has been selected.
t Select the desired setting for the RS-232 interface.
Baudrate: 600 to 19,200 (factory setting: 9600)
Databits: 7 bits or 8 bits (factory setting: 8 bits)
Parity: odd, even, or none (factory setting: odd)
Stopbits: 1 bit or 2 bits (factory setting: 1 bit)
Handshake: software (XON, XOFF), hardware (CTS, RTS), or off
[factory setting: hardware (CTS, RTS)]