Error Messages
Operating Instructions Moisture Analyzer MA160
Calibration/Adjustment Error Messages
“Weight is too light.” A weight that is too light has been placed during
external calibration.
“Weight is too heavy.” A weight that is too heavy has been placed during
external calibration.
“Not able to save
cal./adj. data.”
The memory is full.
Information about Analyzer Maintenance
“Maintenance interval
expired”
The maintenance date set by the
Sartorius Service Center has been exceeded.
USB Device Error Message
“The attached USB device
is not supported.”
A USB device (printer) that is not permitted by
Sartorius has been connected.
Problem
Meaning Remedy
Selected temperature is too high and the
sample is oxidized;
sample is boiling or burning and the
splashes are continuously changing the
weight.
– Reduce drying temperature
– Lay glass-fiber filter onto the
sample
– Reduce the sample quantity or
apply the sample more evenly
– Select a semi-automatic switch-off
criterion or select timer mode
Measurement time too long – Increase temperature
– Reduce sample quantity
– Preheat the analyzer by running a
drying process for 2 or 3 minutes
with an empty sample pan, for
example
Sample loses weight before it is analyzed – Remove the sample pan and apply
the sample outside the sample
chamber
Sample is liquid or pasty – Use glass fiber filter
Sample has only a low moisture content – Increase sample quantity
Insufficient heating output – Clean temperature sensor
Place of installation unstable
(vibrations, etc.)
– Set up the scale in another area