68  Cubis MSE Operating Instructions
Data Interfaces
Data Output
You can define the data output parameter so that output is activated either when 
a print command is received or automatically synchronized with the display or at 
defined intervals (see application programs and autoprint settings)�
  Data Output by Print Command  The print command can be transmitted by pressing P or by a software 
command (EscP)�
  Automatic Data Output  In autoprint mode, data is output to the data interface port without an extra print 
command� You can have synchronized data output automatically at defined display 
update intervals, with or without the stability parameter� The interval time depends 
on the balance operating status and balance type�
If the automatic data output is activated in the Device Configuration, it starts 
immediately after the balance is turned on� You can also configure whether the 
automatic data output can be stopped and started by pressing the P key�
Data Output Formats
You can output the values displayed in the line for measured values and weight 
units with or without an ID code� Configure this output parameter in the Device 
Parameters menu (Menu > Device parameters > Configure data output > Line 
format)�
  Example: Output Without an ID Code  +     253 pcs  16 characters are printed
  Example: Output With an ID Code  Qnt   +     253 pcs  22 characters are printed
Data Output Format with 16 Characters
Display segments that are blank are output as spaces� 
Display values without a decimal point are printed without a decimal point�
The type of character that can be printed depends on the character’s position�
Normal Operation
Position 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16
  + A A A A A A A A A *  E E E CR LF
or  – � � � � � � � � � � * * * 
or    * * * * * * * * * *
*:  Space  CR:  Carriage return
A:  Displayed characters  LF:  Line feed
S:  Characters for the unit