Cubis MSE Operating Instructions  69
Data Interfaces
Special Outputs
Position 1  2  3  4  5  6  7  8  9  10 11 12 13 14 15 16
  * * * * * * *  * * * * * * * CR LF
or    H i g h
or    L o w
or        C a  l  �  E x t  �
*:  Space  High: Overload
Cal� Ext�:  Calibration, external  Low:  Underload
W:  Draft shield status (optional)
I:  Ionizer (optional)
Y,Y,Y =  Draft shield doors
XXX =   Decimal value calculated from binary data:   
Control information for models with normal draft shields
Decimal value   Binary value  Control information
 
1
  Bit0 = 0:  No error/ionizer off
    Bit0 = 1:  Draft shield error/ionizer on
 
2
  Bit1 = 0:  Draft shield motors off
    Bit1 = 1:  Draft shield in motion
 
8
  Bit3 = 0:  Learning function off
    Bit3 = 1:  Learning function on
 
16
  Bit4 = 0:  At least one draft shield door open
    Bit4 = 1:   All draft shield doors closed
 
32
  Bit6 = 0:  Motorized draft shield operation
    Bit6 = 1:  Manual draft shield operation
R,M,L = COO: Right door closed (Closed), Middle and Left doors open (Open)
R,M,L = OCC: Right door open (Open), Middle and Left doors closed (Closed)
Error message
Position 1  2  3  4  5  6  7  8  9  10 11 12 13 14 15 16
       E r r * # # # * * * * CR LF
        A  P  P  �  E  R R
1
)
 * * * * CR LF
        D  I  S  �  E  R R
1
)
 * * * * CR LF
        P R T  �  E R R
1
)
 * * * * CR LF
*:  Space  # # #:  Error code number
1
)  For cause and solution, please refer to “Error Messages”
Example: Output of the weight value + 123,56 g
Position 1  2  3  4  5  6  7  8  9  10 11 12 13 14 15 16
  + * *  * 1 2 3  � 5 6 * g *  * CR LF
  +  *  *  1  2  3  �  5  [  6 ]
1
)  g  *  *  CR LF
Position 1:  Plus +, or minus – or space
Position 2:  Space
Positions 3 – 10:   Weight value with decimal point; leading zeros are output  
as spaces�
Position 11:  Space
Position 12 – 14:  Characters for unit of measure or space
Position 15:  Carriage Return