70  Cubis MSE Operating Instructions
Data Interfaces
Data Output Format with 22 Characters
Here, the data output format with 16 characters is preceded by an ID code  
with a 6 characters� These six characters identify the subsequent value�
Normal Operation
1  2  3  4  5  6  7  8  9  10 11 12 13 14 15 16 17 18 19 20 21 22
K K K K K K + A A A A A A A A A  * E E E CR LF
* * * * * * – � � � � � � � � � * * * *
            * * * * * * * * * *
K:  ID code character
1
)  S:  Unit symbol character
1
)
*:  Space    See Section “Mass Unit Conversion”
A:  Displayed characters  CR:  Carriage return
LF:  Line feed
Example:
 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22
 N  +  1 2 3 � 5 6 * g * * CR LF
 N  +  1 2 3 � 5 [ 6 ]
1
)  g  *  *  CR LF
1
)  Identification of non-verified digits 
Non-verified digits where „e=d“ are automatically marked on the printout by 
square brackets when you select the following setting: Configuration:  
Communication mode: Printer (code 1� 5� 6� 2)� 
 
SBI setting: 
In the SBI setting (code 1� 5� 6� 1), the unverified values are not automatically 
identified� Please take appropriate measures or adjust the settings on the  
peripheral device�
Special Outputs
 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22
 S t a t * * * * * * * * * * * * * * * * CR LF
  H i g h
  L o w
  C a l � E x t �
*:  Space  High: Overload
Cal� Ext�:  Calibration, external  Low:  Underload
Error message
 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22
 S t a t * * * * * E R R * # # # * * * * CR LF
 S t a t * * * * * A P P � E R R 
2
)
  *  *  *  *  CR LF
 S t a t * * * * * D I S � E R R 
2
)
  *  *  *  *  CR LF
 S t a t * * * * * P R T � E R R 
2
)
  *  *  *  *  CR LF
*:  Space  # # #:  Error code number
2
)  For cause and solution, please refer to „Error Messages“