Cubis MSE Operating Instructions 91
Technical Specifications
Model-specific Data
Precision balances
Models MSE623S MSE623P MSE323S
Readability mg 1 1/2/5 1
Weighing capacity g 620 150/300/620 320
Tare range (subtractive) g – 620 – 620 – 320
Repeatability <±mg 0�7 1/2/4 0�7
Linearity <±mg 2 5 2
Corner load (test load [g]) mg 2 (200) 4 (200) 2 (200)
Optimal starting point
of the measuring range* g 0�82 0�82 0�82
Sensitivity drift between +10 to +30°C ±ppm/K 2 2 2
Typical stabilization time s < 0�8 < 0�8 < 0�8
Typical measurement time s < 1 < 1 < 1
External standard calibration value g 500 (E2) 500 (F1) 200 (E2)
(min� accuracy class)
Display result
(depending on the set filter level) 0�1 – 0�4
Weighing pan size (W + D) mm 140 + 140
Weighing chamber height
(draft shield DE) mm 172
Protection Protected against dust and water
* = According to USP (United States Pharmacopeia) section 41, the optimal measuring range is defined from 820 d to the max� weighing
range� Typically the value can be higher depending on the setup location and the ambient conditions�
Models MSE14202S MSE14202P MSE10202S MSE8202S
Readability mg 10 10/20/50 10 10
Weighing capacity g 14,200 3,500/7,000/14,200 10,200 8,200
Tare range (subtractive) g –14,200 –14,200 –10,200 –8,200
Repeatability <±mg 10 10/20/40 7 7
Linearity <±mg 30 50 20 20
Corner load (test load [g]) mg 20 (5,000) 40 (5,000) 20 (5,000) 20 (5,000)
Optimal starting point
of the measuring range* g 8�2 8�2 8�2 8�2
Sensitivity drift between +10 to +30°C ±ppm/K 1�5 1�5 2 2
Typical stabilization time s < 1 < 1 < 1 < 1
Typical measurement time s < 1�5 < 1�5 < 1�5 < 1�5
External standard calibration value kg 10 (E2) 10 (E2) 10 (E2) 5 (E2)
(min� accuracy class)
Display result
(depending on the set filter level) 0�1 – 0�4
Weighing pan size (W + D) mm 206 + 206
IP protection
(according to EN 60529/IEC 60529) IP54
* = According to USP (United States Pharmacopeia) section 41, the optimal measuring range is defined from 820 d to the max� weighing
range� Typically the value can be higher depending on the setup location and the ambient conditions�