92 Cubis MSE Operating Instructions
Technical Specifications
Model-specific Data
Precision balances
Models MSE6202S MSE6202P MSE5202S MSE4202S
Readability mg 10 10/20/50 10 10
Weighing capacity g 6,200 1,500/3,000/ 5,200 4,200
6,200
Tare range (subtractive) g – 6,200 – 6,200 – 5,200 – 4,200
Repeatability <±mg 7 7/20/40 6 7
Linearity <±mg 20 50 10 20
Corner load (test load [g]) mg 20 (2,000) 50 (2,000) 10 (2,000) 30 (2,000)
Optimal starting point
of the measuring range* g 8�2 8�2 8�2 8�2
Sensitivity drift between +10 to +30°C ±ppm/K 2 2 2 2
Typical stabilization time s < 1 < 1 < 0�8 < 0�8
Typical measurement time s < 1�5 < 1�5 < 1 < 1
External standard calibration value kg 5 (E2) 5 (F1) 5 4 (E2)
(min� accuracy class)
Display result
(depending on the set filter level) 0�1 – 0�4
Weighing pan size (W + D) mm 206 + 206 206 + 206 140 + 140 206 + 206
IP protection
(according to EN 60529/IEC 60529) IP54
* = According to USP (United States Pharmacopeia) section 41, the optimal measuring range is defined from 820 d to the max� weighing
range� Typically the value can be higher depending on the setup location and the ambient conditions�
Models MSE2202S MSE1202S MSE70201S MSE36201S MSE36201P
Readability mg 10 10 100 100 100/1,000
Weighing capacity g 2,200 1,200 70,200 36,200 10,200/36,200
Tare range (subtractive) g – 2,200 – 1,200 – 70,200 – 36,200 – 36,200
Repeatability <±mg 7 7 70 70 70/500
Linearity <±mg 20 20 500 200 200
Corner load (test load [g]) mg 20 (1,000) 20 (500) 500 (20,000) 300 (10,000) 300 (10,000)
Optimal starting point
of the measuring range* g 8�2 8�2 82 82 82
Sensitivity drift between +10 to +30°C ±ppm/K 2 2 3 2 2
Typical stabilization time s < 0�8 < 0�8 < 1 < 1 < 1
Typical measurement time s < 1 < 1 < 1�5 < 1�5 < 1�5
External standard calibration value kg 2 (F1) 1 (F1) 50 (F1) 20 (F1) 20 (F1)
(min� accuracy class)
Display result
(depending on the set filter level) 0�1 – 0�4
Weighing pan size (W + D) mm 206 + 206 206 + 206 400 + 300 400 + 300 400 + 300
IP protection
(according to EN 60529/IEC 60529) IP54
* = According to USP (United States Pharmacopeia) section 41, the optimal measuring range is defined from 820 d to the max� weighing
range� Typically the value can be higher depending on the setup location and the ambient conditions�