Cubis MSE Operating Instructions 93
Technical Specifications
Model-specific Data
Precision balances
Models MSE20201S MSE12201S MSE8201S MSE5201S
Readability mg 100 100 100 100
Weighing capacity g 20,200 12,200 8,200 5,200
Tare range (subtractive) g – 20,200 – 12,200 – 8,200 – 5,200
Repeatability <±mg 70 50 50 50
Linearity <±mg 200 100 100 100
Corner load (test load [g]) mg 300 (5,000) 200 (5,000) 200 (5,000) 200 (2,000)
Optimal starting point
of the measuring range* g 82 82 82 82
Sensitivity drift between +10 to +30°C ±ppm/K 2 4 4 4
Typical stabilization time s < 1 < 0�8 < 0�8 < 0�8
Typical measurement time s < 1�5 < 1 < 1 < 1
External standard calibration value kg 20 (F1) 10 (F1) 5 (F2) 5 (F2)
(min� accuracy class)
Display result
(depending on the set filter level) 0�1 – 0�4
Weighing pan size (W + D) mm 206 + 206
IP protection
(according to EN 60529/IEC 60529) IP54
* = According to USP (United States Pharmacopeia) section 41, the optimal measuring range is defined from 820 d to the max� weighing
range� Typically the value can be higher depending on the setup location and the ambient conditions�
Models MSE70200S MSE36200S
Readability mg 1,000 1,000
Weighing capacity g 70,200 36,200
Tare range (subtractive) g – 70,200 – 36,200
Repeatability <±mg 500 500
Linearity <±mg 1,000 1,000
Corner load (test load [g]) mg 1,000 (20,000) 1,000 (10,000)
Optimal starting point
of the measuring range* g 820 820
Sensitivity drift between +10 to +30°C ±ppm/K 3 2
Typical stabilization time s < 0�8 < 0�8
Typical measurement time s < 1 < 1
External standard calibration value kg 50 (F1) 20 (F1)
(min� accuracy class)
Display result
(depending on the set filter level) 0�1 – 0�4
Weighing pan size (W + D) mm 400 + 300
IP protection
(according to EN 60529/IEC 60529) IP54
* = According to USP (United States Pharmacopeia) section 41, the optimal measuring range is defined from 820 d to the max� weighing
range� Typically the value can be higher depending on the setup location and the ambient conditions�