Technical Specifications
Model-specific Data
Verified models with EC Type Approval Certificate: Precision balances
Models MSE5203S-0CE MSE5203P-0CE MSE3203S-0CE MSE3203P-0CE
Accuracy class* k k k k
For verified models: EC Type-Approval Certificate D09-09-015, Type: MSX
Scale interval d* mg 1 1/2/5 1 1/10
Weighing capacity max* g 5,200 1,200/2,400/5,200 3,200 1,010/3,200
Calibration value e* mg 10 10 10 10
Min� load min* mg 100 100 100 100
Tare equalization range (subtractive) < 100% from max� weighing capacity
Application range according to DIR* g 0�1 – 5,200 0�1 – 5,200 0�1 – 3,200 0�1 – 3,200
Optimal starting point
of the measuring range** g 0�82 0�82 0�82 0�82
Typical stabilization time s < 1 < 1 < 1 < 1
Typical measurement time s < 2 < 2 < 2 < 1�5
External standard calibration value g 5,000 (E2 5,000 (E2 2,000 (E2 2,000 (E2)
(min� accuracy class)
Application range (temperature) With “isoCAL” function: +5 ��� +40 °C
Without “isoCAL” function: +15 ��� +25 °C
Display result
(depending on the set filter level) 0�1 – 0�4
Weighing pan size (W + D) mm 140 + 140
Weighing chamber height (draft shield DE)
mm 172
Protection Protected against dust and water
Models MSE2203S-0CE MSE2203P-0CE MSE1203S-0CE
Accuracy class* k k k
For verified models: EC Type-Approval Certificate D09-09-015, Type: MSX
Scale interval d* mg 1 1/10 1
Weighing capacity max* g 2,200 1,010/2,200 1,200
Calibration value e* mg 10 10 10
Min� load min* mg 100 100 100
Tare equalization range (subtractive) < 100% from max� weighing capacity
Application range according to DIR* g 0�1 – 2,200 0�1 – 2,200 0�1 – 1,200
Optimal starting point
of the measuring range** g 0�82 0�82 0�82
Typical stabilization time s < 1 < 1 < 1
Typical measurement time s < 1�5 < 1�5 < 1�5
External standard calibration value g 2,000 (E2) 1,000 (E2) 1,000 (E2)
(min� accuracy class)
Application range (temperature) With “isoCAL” function: +5 ��� +40 °C
Without “isoCAL” function: +15 ��� +25 °C
Display result
(depending on the set filter level) 0�1 – 0�4
Weighing pan size (W + D) mm 140 + 140
Weighing chamber height (draft shield DE)
mm 172
Protection Protected against dust and water
* DIR = Directive 2014/31/EU on non-automatic weighing instruments used within the European Economic Area
** = According to USP (United States Pharmacopeia) section 41, the optimal measuring range is defined from 820 d to the max� weighing
range� Typically the value can be higher depending on the setup location and the ambient conditions�
96 Cubis MSE Operating Instructions