Technical Specifications
Model-specific Data
Verified models with EC Type Approval Certificate: Precision balances
Models MSE6202S- MSE6202P- MSE5202S- MSE4202S-
0CE 0CE 0CE 0CE
Accuracy class* K K K K
For verified models: EC Type-Approval Certificate D09-09-015, Type: MSX
Scale interval d* g 0�01 0�01/0�02/0�05 0�01 0�01
Weighing capacity max* g 6,200 1,500/3,000/6,200 5,200 4,200
Calibration value e* g 0�1 0�1 0�1 0�1
Min� load min* g 0�5 0�5 0�5 0�5
Tare equalization range (subtractive) < 100% from max� weighing capacity
Application range according to DIR* g 0�5 – 6,200 0�5 – 6,200 0�5 – 5,200 0�5 – 4,200
Optimal starting point
of the measuring range** g 8�2 8�2 8�2 8�2
Typical stabilization time s < 1 < 1 < 0�8 < 0�8
Typical measurement time s < 1�5 < 1�5 < 1 < 1
Application range (temperature):
With “isoCAL” function +5 to +40 °C +5 to +40 °C +5 to +40 °C
Without “isoCAL” function +10 to +30 °C +10 to +30 °C +10 to +30 °C
Display result
(depending on the set filter level) 0�1 – 0�4
Weighing pan size (W + D) mm 206 + 206 206 + 206 140 + 140 206 + 206
IP protection
(according to EN 60529/IEC 60529) IP54
Models MSE2202S- MSE1202S- MSE36201S- MSE36201P- MSE20201S-
0CE 0CE 0CE 0CE 0CE
Accuracy class* K K K K K
For verified models: EC Type-Approval Certificate D09-09-015, Type: MSX
Scale interval d* mg 10 10 100 100/1,000 100
Weighing capacity max* g 2,200 1,200 36,200 10,200/36,200 20,200
Calibration value e* g 0�1 0�1 1 1 1
Min� load min* g 0�5 0�5 5 5 5
Tare equalization range (subtractive) < 100% from max� weighing capacity
Application range according to DIR* g 0�5 – 2,200 0�5 – 1,200 5 – 36,200 5 – 36,200 5 – 20,000
Optimal starting point
of the measuring range** g 8�2 8�2 82 82 82
Typical stabilization time s < 0�8 < 0�8 < 1�5 < 1�5 < 1�5
Typical measurement time s < 1 < 1 < 2 < 2 < 2
Application range (temperature) With “isoCAL” function: +5 to +40 °C | Without “isoCAL” function: +10 to +30 °C
Display result
(depending on the set filter level) 0�1 0�4
Weighing pan dimensions (W + D) mm 206 + 206 206 + 206 400 + 300 400 + 300 400 + 300
IP protection
(according to EN 60529/IEC 60529) IP54
* DIR = Directive 2014/31/EU on non-automatic weighing instruments used within the European Economic Area
** = According to USP (United States Pharmacopeia) section 41, the optimal measuring range is defined from 820 d to the max� weighing
range� Typically the value can be higher depending on the setup location and the ambient conditions�
98 Cubis MSE Operating Instructions

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sartorius MSE3.6P and is the answer not in the manual?

Sartorius MSE3.6P Specifications

General IconGeneral
BrandSartorius
ModelMSE3.6P
CategoryScales
LanguageEnglish

Related product manuals