Service Manual Signum
Working with the Service Mode
Calibration/Adjustment Functions
Important note:
When using a verified weighing platform, open the cover on the back,
left-hand side of the indicator housing. Then slide the menu access
switch to the right (»free« setting).
Activate the service mode (see page 15).
1-9-1 External Calibration/Adjustment with Default Weights
Select weighing platform »WP-1«.
Press the ) key to access the number menu.
Select menu »1-9-1« (External calibration/adjustment with default weight) by
pressing the ) k ) key repeatedly and then pressing the ) key briefly
to confirm.
If this menu has already been selected, switch the indicator off and then on
again to exit the service mode.
display_wps1_serv.eps
display_serv_01.eps
menu_191.eps