Service Manual Signum
Working with the Service Mode
1-9-6 External Linearization with Default Weights
Activate the service mode (see page 15).
Select weighing platform »WP1«.
Press the ) key to access the number menu.
Press the ) and k)) keys repeatedly to select menu code »1-9-6« (Ext.
Linearization with Default Weight). Then briefly press the ) key to confirm.
Press and hold the ) key to store the selected menu and return the scale to
the weighing mode.
If this menu has already been selected, switch the indicator off and on again to
exit the service mode.
After taring or zeroing the display of the scale, press the J key until »L.EXT.
DEF« is displayed.