Service Manual Signum
Working with the Service Mode
Press the J key to store the 3
rd
linearization weight. The 4
th
required weight is
now displayed.
Place the required weight on the scale. After a brief pause, the difference
between the current value and the value from the last adjustment (calibration)
procedure is displayed
Press the J key to store the 4
th
linearization weight. The scale now prompts a
zero point (remove all weights from the scale).
Once the zero point is stored, the scale will automatically return to the weighing
mode.
*)
Weights displayed depend on the scale model.
3)
Only displayed if menu code »1-10-2« (Calibration with manual adjustment) has been selected.