24
Service Manual Signum 
Working with the Service Mode
–  After approx. 2 seconds, the 1
st
 required linearization weight (defined weight) 
is displayed. You can enter the weights for linearization via menus »1-18-2« 
through »1-18-5«
2)
.
–  Place the required weight on the scale.  After a brief pause, the difference 
between the current value and the value from the last adjustment (calibration) 
procedure is displayed. 
–  Press the ( key if you need to interrupt the calibration/adjustment procedure.
–  Press the J key to store the 1
st
 linearization weight.  Next, the 2
nd
 required 
weight is displayed.
–  Place the required weight on the scale. After a brief pause, the difference 
between the current value and the value from the last adjustment (calibration) 
procedure is displayed.
–  Press the J key to store the 2
nd
 linearization weight.  Now the 3
rd
 required 
weight is displayed.
–  Place the required weight on the scale. After a brief pause, the difference 
between the current value and the value from the last adjustment (calibration) 
procedure is displayed.
*)
 Weights displayed depend on the scale model 
2)
 Only possible in the service mode 
3)
 Only displayed if menu code »1-10-2« (Calibration with manual adjustment) has been selected.
2.5kg.eps
display_serv_02.eps
5kg.eps
display_serv_03.eps
7.5kg.eps
display_serv_04.eps
*
*
*
 
3)
 
3)
 
3)