32
Service Manual Signum
Working with the Service Mode
1-18 Entering Calibration/Linearization Weights
1-18-1 External User-Defined Calibration Weight (service mode not required)
Activate the service mode (see page 15).
Select weighing platform »WP1«.
Press the ) key to access the number menu.
Select menu code »1-18-1 « (Ext. user – defined calibration weight) by pressing
the ) key and the k ) key repeatedly).
Access the input mode for the user-defined weight by briefly pressing the )
key. The first number blinks.
Enter the user-defined weight (10,000 kg in this example); press the following
keys in order: ) k ) p ) ) ) k ). Now the last number
entered blinks.
Press the ) key again to store the user-defined weight.
display_wps1_serv.eps
display_serv_01.eps
menu_1181.eps
user1.eps
user2.eps
menu_1181.eps