Internal Calibration
SIWS models:
Verified scale or scale equipped with
Option E7
In the operating menu under setup:
wp-1: 1.9., the “Internal calibration"
option (
menu code 1.9.4) must be set.
setup
wp-1
1. 9. Calibration,
adjustment
1. 9. 1 * External cal./
adjustment;
default weight
1. 9. 3 External cal./
adjustment;
weight can be
selected under
menu item 1.18. 1)
1. 9. 4 Internal calibra-
tion/adjustment
(only on models
with a built-in
motorized calibra-
tion weight)
1. 9. 8 Set preload
1. 9. 9 Clear preload
1. 9.10 J key blocked
1.10. Calibration/
adjustment
sequence
1.10. 1 Calibration with
automatic
adjustment
1.10. 2 * Calibration with
adjustment
triggered manually
* = Factory settings
The built-in motorized calibration
weight is applied and removed auto-
matically for internal calibration.
The adjustment procedure is as follows:
§ Select calibration/adjustment:
Press the J key
> The built-in weight is applied
automatically
> The scale is adjusted
> If menu code 1.10.1 is selected
under
Setup: wp-1: 1.10., the scale is
adjusted automatically
> If menu code 1.10.2 is selected under
Setup: wp-1: 1.10., the internal cali-
bration routine can be stopped at this
point, before adjustment is carried out
> The built-in calibration weight is
unloaded from the weighing system
> ISO/GMP-compliant printout:
see page 80
Setting the Preload
Notes on Settings
!The preload can be set only when the
menu access switch is open.
The “set preload" function must be
assigned to the J key (menu item
1.9.8).
!After setting the preload, close the
menu access switch and reset the J
key to its previous function (e.g., exter-
nal calibration/adjustment with user-
defined weights) under menu item 1.9.
Clearing the Preload
Notes on Settings
!The preload can be cleared only when
the menu access switch is open.
The “clear preload" function must be
assigned to the J key (menu item
1.9.9).
!After clearing the preload, close the
menu access switch and reset the J
key to its previous function (e.g., exter-
nal calibration/adjustment with user-
defined weights) under menu item 1.9.

Table of Contents

Other manuals for Sartorius Signum 3

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sartorius Signum 3 and is the answer not in the manual?

Sartorius Signum 3 Specifications

General IconGeneral
BrandSartorius
ModelSignum 3
CategoryScales
LanguageEnglish

Related product manuals