Example: 1
Checkweighing samples with a target weight of 1250 g and a tolerance range from –10 g to +30 g
Settings (changes in the factory settings required for this example):
Application: Checkweighing toward zero (menu item 4.7.2)
Setup: Printout:
PrtProt: 7.6.; then select menu line item of your choice
(see “Configuration” for options)
O
1 Enter the initial target and
tolerance limit values
2 Place a sample equal to the
target weight (in this example,
1250 g) on the platform
O
3 Save target value
1240
4 Enter value for lower limit
(in this example, 1240 g)
O
5 Save value for the lower limit
1280
6 Enter value for the upper limit
(in this example, 1280 g)
O
Save value for upper limit
7 Weigh samples
p
8 Print the results
Note: If automatic printout
of results is enabled, you do
not need to press the p key;
the results are printed auto-
matically.
Setp + 1.250 kg Target
Min + 1.240 kg Minimum
Max + 1.280 kg Maximum
G# + 1.256 kg Gross weight
T + 0.000 kg Tare weight
N + 1.256 kg Net weight
Lim + 0.48 % Percentage of deviation from
target*
W.Diff+ 0.006 kg Absolute deviation from target
-------------------
* Only in “Tolerance limits”
display mode:
If the weight is lighter than
the target, the display shows:
LL
If the weight is heavier than
the target, the display shows:
HH
Application 2: Checkweighing toward Zero
67