Example
Determining sample weights while monitoring the minimum sample quantity (in this example, SQmin = 100 g)
Settings (changes in the factory settings required for this example):
Setup: SQmin: Display
Step Key (or action) Display/printout
1 Switch on the scale and configure e
settings as indicated above
2 Place the container for the )
sample on the scale and tare
3 Measure the weight of a sample Place sample on scale
(in this example:
minimum sample quantity not reached)
4 Print weight p | N + 90.0 ! |
5 Measure the weight of another sample Place sample on scale
(in this example:
minimum sample quantity exceeded)
6 Print weight p | N + 110.0 g |
7 Show value of minimum sample k
quantity for 4 seconds
8 Weigh other samples as desired